Determination of Manganese (II) using Flow Injection Analysis with Chemiluminescence detection

Main Article Content

K. M. Omer

Abstract

The determination of manganese (II) using flow injection analysis with chemiluminescence detection was investigated. Mn2+ in sample solutions injected into a carrier stream of sodium bismuthate (NaBiO3) were oxidised to form MnO4- ions which were capable of producing luminescence after reaction with luminol/KOH in a flow cell. The linear range of the system is from 20 to 80 mg/L with a detection limit 8 mg/L. The proposed system is suitable for determination of Mn2+ in steel alloys after dissolution, filtration and dilution at a rate of approximately 60 samples per hour with a relative standard deviation (RSD)1.2%. Statistical comparison between the proposed system and standard spectrophotometric method revealed that there is no significant difference between the two methods.

Article Details

How to Cite
Determination of Manganese (II) using Flow Injection Analysis with Chemiluminescence detection. (2019). Ibn AL-Haitham Journal For Pure and Applied Sciences, 18(3), 111-120. https://jih.uobaghdad.edu.iq/index.php/j/article/view/2228
Section
Chemistry

How to Cite

Determination of Manganese (II) using Flow Injection Analysis with Chemiluminescence detection. (2019). Ibn AL-Haitham Journal For Pure and Applied Sciences, 18(3), 111-120. https://jih.uobaghdad.edu.iq/index.php/j/article/view/2228