Determination of Manganese (II) using Flow Injection Analysis with Chemiluminescence detection

Authors

  • K. M. Omer

Keywords:

Manganese, Flow injection analysis, Chemiluminescence

Abstract

The determination of manganese (II) using flow injection analysis with chemiluminescence detection was investigated. Mn2+ in sample solutions injected into a carrier stream of sodium bismuthate (NaBiO3) were oxidised to form MnO4- ions which were capable of producing luminescence after reaction with luminol/KOH in a flow cell. The linear range of the system is from 20 to 80 mg/L with a detection limit 8 mg/L. The proposed system is suitable for determination of Mn2+ in steel alloys after dissolution, filtration and dilution at a rate of approximately 60 samples per hour with a relative standard deviation (RSD)1.2%. Statistical comparison between the proposed system and standard spectrophotometric method revealed that there is no significant difference between the two methods.

Downloads

Published

17-Jun-2019

Issue

Section

Chemistry

Publication Dates